8 Test and Maintenance Software
8-
15
Figure 8-14 Enable Modifying Parameters 2
NOTE
1、 The configuration modification is only effective to selected
unit.If the number of the units is more than 1, modify configuration
of each unit respectively.
2 If the main board is in the status of parameter write
protection, you can still use the method to modify parameter
configuration, but the parameters will be restored to the previous
one after restarting.
3 If the main board is not in the status of parameter write
protection, the modification will be validated perpetually.
8.3.2 Detailed Operations
The operation units related to the parameters include Inquire, Configure, Save, Read
and Configure All. Enable the parameter modification before configuring the parameters,
The detailed operations are described in the section 8.3.1
Before performing the operation, select the target unit from the drop-down list of “Unit
Name” or “Speed Name”. If a unit from the Unit Name list and a motor from the Speed
Name list are both selected, the system can only identify the last selected one.