8-16 8 Test and Maintenance Software
8.3.2.1 Query
Select the target unit and click Query to inquire about the parameters of the
corresponding units and the motors.
8.3.2.2 Config
1. Inquire or read the parameters of the selected unit.
2. Modify the parameter
3. Enable the parameters to be selectable and then click Config.
4. The parameter configuration will be completed, only when the Config
command frame (the third byte of the communication frame is 01), Reply
frame (the third byte of the communication frame is 02) and the result frame
(the third byte of the communication frame is 03) show up.
5. After parameter configuration, click Query to confirm.
The parameter configuration will be successful only when the parameter modification of
the target unit is enabled. Otherwise, the warning of parameter modification protection will
show up at the bottom of the screen.
For the units in the Unit Name drop-down list, if the main board is in parameter write
protection status, the new parameters will be validated after configuration; but the system
will restore the previous parameters after restarting. If the modification function is enabled,
the new parameters will be perpetually validated after configuration.
For the motors in the Speed Name list, the configuration modification will only be validated
after configuration. After the analyzing unit is shutdown, the motor parameters will be
stored to the original ones.
8.3.2.3 Save
Click Save to back-up the current parameters of the target unit in a new file after the
operations of Inquire, Configure, Read and Configure All.
8.3.2.4 Load
1)Select a target unit.
2)Click Read and then select a parameter file in the pop-up dialog box. If the selected file
is not belong to the target unit, the system will automatically switch the current unit to the
unit specified by the parameter file.
3) Click Config All to refresh all the parameters of the target unit.
8.3.2.5 Config All
1)
Select a target unit.
2) Inquire or read the parameters of the selected unit, or modify the parameter if needed.
3) Select Confiure All to complete the configuration with all current parameters to the
target unit.