7-36 7 Service and Maitenance
6 Place the syringe on the holder. Install space bars and fix retaining
screws.
NOTE
The upper edge of the upper space bar must reach the
seventh line of the scale on the syringe.
When fixing the retaining screws, be sure to tighten them
alternately with equilibrium force.
7 Screw (clockwise) the lower retaining screw until secure.
7.6.9 Replacing Dust Screen
Replace the dust screen of poor ventilation after long time use. Please refer to 7.5.1
Washing Dust Screen for details.
7.6.10 Replacing Waste Tubing
If the waste tubing cannot discharge waste smoonthly after being used for a long time,
replace it with the specified one.
7.7 Maintaining ISE Module (Optional)
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
Dispose of the waste part in accordance with your local or national
guidelines for biohazard waste disposal.
CAUTION
Please use Mindray-recommended consumables. Other
consumables may decrease the system performance.
NOTE
Run ISE calibration after the replacement of any following
components.
7.7.1 Replace Reagent Pack
1 Place the Power to OFF.
2 Open the ISE unit door.
3 Remove and install a new reagent module.
4 Enter the ISE screen of the Maintenance of the system
software.