9 Troubleshooting 9-49
100701250005
ISE unit result error:
Instruction sending
failed
Serial port cable between ISE
module and the main board
not connected or serial port
cable error
Check the connection between the ISE module and the
main board. Check whether the serial port cable is
wrong.
100701260005
ISE unit result error:
Main unit does not
receive response from
ISE unit
Serial port cable between ISE
module and the main board
not connected or serial port
cable error
Check the connection between the ISE module and the
main board. Check whether the serial port cable is
wrong.
100701270005
ISE unit result error:
Main unit does not
receive results from ISE
unit
Serial port cable between ISE
module and the main board
not connected or serial port
cable error
Check the connection between the ISE module and the
main board. Check whether the serial port cable is
wrong.
9.4.7 Other Units Failures
Error Code Level Error Message
Reason Corrective Measure
A1401 11 Undefined
The automatically generated
error code not in the edited
range.
Upgrade the operating software or contact
the service personnel.