9 Troubleshooting 9-23
9.4.4 Temperature Unit
Error Code Error Message Reason Corrective Measure
100660010000
Temperature unit result
error: Command error
Command sending error
or wrong command
information
Reset the Power, restart the operating software,and retry
this command. If the error remains, please contact the
R&D engineer.
100660020000
Temperature unit result
error: Self-check error
Self-check error
Shut down the Power, restart the operating software, and
retry this command. If the error remains, turn off the
analyzing unit, check the temperature sensor and the
cable.
100660030000
Temperature unit result
error: Mechanical
resetting error
Mechanical resetting error
Reset the mechanical parts by using Alignment. If the
error remains, please contact the R&D engineer.
100660040000
Temperature unit result
error: Status error.
Self-check
Self-checking; Other
operations not supported
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting.
100660040010
Temperature unit result
error: Status error. Error
Temperature unit error;
current operations not
supported
Execute failure recovery. If it repeats continuously for
several times, please contact the R&D engineer.
100660040020
Temperature unit result
error: Status error.
Waiting for handshake
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting. If it repeats for several times, please
contact the R&D engineer.
100660050000
Temperature unit result
error: Unit busy. No
response
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting. If it repeats for several times, please
contact the developer.
100660070000
Temperature unit result
error: Wrong searching
parameter
Wrong system operation
parameter
Input the right searching parameter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BS-130 and is the answer not in the manual?

Mindray BS-130 Specifications

General IconGeneral
BrandMindray
ModelBS-130
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals