9 Troubleshooting
9-22
100650150025
Reaction unit result
error: Photoelectric
error. Signal collection
busy
Running photoelectric
collection. Other
operations not supported.
Run other operations, after photoelectric collection is
finished.
100650150035
Reaction unit result
error: Photoelectric
error. Lamp On/Off
actions are opposite
Lamp On/Off actions are
opposite
Adjust the lamp by using Alignment. If the error remains,
contact the R&D engineer.
100650150045
Reaction unit result
error: Photoelectric
measurement error.
Filter home position lost.
Filter wheel step missing
or error
Shut down the Power, restart the operating software, and
retry this command. If the error remains, contact the R&D
engineer.
100650150055
Reaction unit result
error: Photoelectric
measurement error.
Timeout.
Other units connection
error
Check the status of the filter wheel and the connection of
the optical module. Reset the Power and rerun the
command. If the error remains, contact the R&D engineer.
100650190005
Reaction unit result
error: Parameter modify
protection
Unit parameter write
protection. Parameter
configuration forbidden
Configure parameter after parameter configuration is
permitted.
100651250005
Reaction unit result
error: Main unit sending
error
/
Shut down Power, restart the operating software, and
retry this command. If the error remains, check the cable
and the board.
100651270005
Reaction unit result
error: Main unit does not
receive the reaction unit
result
/
Shut down the Power, restart the operating software, and
retry this command. If the error remains, check the cable
and the board.