9 Troubleshooting 9-19
100680220095
Sample unit result error:
Sample probe vertical
movement error. Sample
disk or reaction disk is
rotating
Vertical movement not
permitted
Check the status of the probe and the disk. Execute
mechanical resetting.
100680220105
Sample unit result error:
Sample probe vertical
movement error.
Disabled
Vertical movement not
permitted
Check the status of the probe and the disk. Execute
mechanical resetting.
100680260005
Sample unit result error:
Parameter write
protection of this unit
Unit parameter write
protection. Parameter
configuration forbidden
Configure parameter after parameter configuration is
permitted.
100681250005
Sample unit result error:
Main unit sending error
/
Shut down Power, restart the operating software, and retry
this command. If the error remains, check the cable and the
board.
100681270005
Sample unit result error:
Main unit does not
receive the sample unit
result
/
Shut down Power, restart the operating software, and retry
this command. If the error remains, check the cable and the
board.
9.4.3 Reaction Disk Unit
Error Code Error Message Reason Corrective Measure
100650010005
Reaction unit result
error: Command error
Commamd sending error
or wrong command
information
Reset Power and restart the operating software. Retry this
command. If the error remains, please contact the R&D
engineer.
100650020005
Reaction unit result
error: Self-check error
Self-check error
Reset Power and restart the operating software. Retry this
command. If the error remains, please contact the R&D
engineer.