9 Troubleshooting
9-18
100680220035
Sample unit result error:
Sample probe vertical
movement error. Trying
to move away from the
sample disk limit
position
Sample probe vertical
movement error. Trying to
move away from the
sample disk limit position
Execute mechanical resetting. If the error remains, check
the status of the sensor.
100680220045
Sample unit result error:
Sample probe vertical
movement error. Trying
to move away from the
washing limit position
Sample probe vertical
movement error. Trying to
move away from the
washing limit position
Execute mechanical resetting. If the error remains, check
the status of the sensor.
100680220055
Sample unit result error:
Sample probe vertical
movement error. Trying
to move away from the
reaction disk limit
position
Sample probe vertical
movement error. Trying to
move away from the
reactiondisk limit position
Execute mechanical resetting. If the error remains, check
the status of the sensor.
100680220065
Sample unit result error:
Sample probe vertical
movement error. No
liquid surface detected
Insufficient sample or
reagent volume
After pausing the sample prbobe, check whether the
sample/reagent is on the right position. If yes, check
whether the level detection board is wrong.
100680220075
Sample unit result error:
Sample probe vertical
movement error.
Collision
Sample probe vertical
movement error. Collision
After pausing the sample probe, check whether the vertical
movement of the sample probe is blocked. If not, check the
status of the sample probe vertical collision protection
sensor and the blocking plate. Execute mechanical
resetting.
100680220085
Sample unit result error:
Sample probe vertical
movement error. Step
missing
Sample probe vertical
movement error. Step
missing
Check the vertical movement motor and the sensor. Check
whether the vertical movement is blocked. Execute
mechanical resetting.