9 Troubleshooting                                                                                    9-11 
400001020001
Test  result  error: 
Calibration SD of … too 
large 
Calibration SD too high 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001030001
Test  result  error: 
Difference  between 
calibration  coefficients 
of … too large 
Something  wrong  with  the 
calibrator and reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001040001
Test  result  error: 
Calibration  related 
coefficients of … too low
Something  wrong  with  the 
calibrator and reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001050001
Test  result  error: 
Incomplete  repeated 
calibration data of … 
Something  wrong  with  the 
calibrator and reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001060001
Test  result  error: 
Calibration  curve  of  … 
not monotonic 
Something  wrong  with  the 
calibrator and reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001070001
Test  result  error: 
Concentration  of  … 
calculation failed 
Something  wrong  with  the 
calibrator and reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001080001
Test  result  error: 
Incomplete  test  result 
of … 
Rerun  calibration  not 
completed.  or  insufficient 
reagent and calibrator. 
Refill the reagent and calibrator; recalibrate. 
400001090001
Test result error: Error of 
repeated … too large 
Something  wrong  with  the 
calibrator and reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators  and  calibration  rule.  Recalibrate.  If  the  error 
remains, check the optical system. 
400001100000
Test result error: QC out 
of control 
Control  performance 
degraded; something wrong 
with the reagent 
Replace  the  reagent  and  calibrators.  Confirm  the 
calibrators and calibration rule.Rerun.If the error remains, 
check the optical system. 
400001140006
Operating  software 
error:  Detergent 
exhausted,  or  invalid 
sensor 
Wash  solution  used  up  or 
sensor invalidated 
Check the DI tank. If there is insufficient DI water, refill it 
and execute failure recovery. 
400001150006
Operating  software 
error: Waste full 
Waste tank full 
Check the  waste tank.  If it  is  full,  empty it  and  execute 
failure recovery. 
400001350000
Test  result  error:  Blank 
response of … too low 
Blank reponse too low  Check the cuvette, reagent and diluent.