7 Service and Maitenance 7-19
WARNING
Do not touch the lamp before it cools down, or you may get burned.Do
not proceed with this procedure until they have cooled down.
2. See Figure 7-1, Unsrew the two screws on the table panel 1 and remove it.
Figure 7-1 Remove the Table Panel
3. Loose the retaining screw manually for about 5-7 circles (do not get the screw
out).
Figure 7-2 The Vew after Table Panel is Removed
4. Use your finger to pinch the porcelain socket and pull it out for 3mm, and then
rotate 45 degrees counterclockwise, to pull the lamp assembly out completely.