9 Troubleshooting
9-34
10070001SEA5
ISE unit result error:
Serum cycle error. Air in
calibrant A
Calibrator A bottle empty;
tube not connected;
pump A not working;
tube blocked, broken or
twisted;
fibrin and salt in the tube;
air bubble detector error;
waste pump error
Wrong electrode installation. Check the pressing plate
spring and the sealing to ensure all the electrodes and
the O-ring are in the right place.
Wash the module by using <CLEN > program. Remove
the module and then wash it. Reinstall the sensor.
Replace air bubble detection sensor;
replace the waste pump;
replace the calibrator A and recalibrate;
reconnect or replace the tube;
check the connection of the motor;
replace the pump box; replace the motor;
replace the tube.
10070001SEF5
ISE unit result error:
Serum cycle error. No
flow
Calibrator A and B bottle
empty; wrong electrode
installation
Replace the calibrator A and B, then recalibrate; check
the installation of the electrodes.
10070001SES5
ISE unit result error:
Serum cycle error. Air in
sample
Insufficient sample volume Rerun after refilling.