9 Troubleshooting 9-41
100700050085
ISE unit result error: Cl
electrode voltage noise
(Cal A in calib mode, Cal
B in urine mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
1007000500A5
ISE unit result error: Cl,
Na electrodes voltage
noise (Cal A in calib
mode, Cal B in urine
mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
1007000500C5
ISE unit result error: Cl,
K electrodes voltage
noise (Cal A in calib
mode, Cal B in urine
mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
1007000500E5
ISE unit result error: Cl,
K, Na electrodes voltage
noise (Cal A in calib
mode, Cal B in urine
mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.