4 Units Description 
4-41
4.8.3.1  ISE Module 
Figure 4-46    ISE Module 
 
The dismounting/mounting steps are as follows (see Figure 4-46 ). 
Dismounting steps: 
1.  Remove the table panel and the ISE unit door. 
2.  Unplug the draining tube connected to the ISE module. 
3.  Loosen the screw and remove the ISE shielding cover. 
4.  Loosen  the  four  M3X8  cross  pan  head  screws  retained  on  the  ISE  shield 
housing and remove the ISE module. 
Mounting steps: 
1.  Install the ISE module on the ISE shield housing by tightening the four M3X8 
cross pan head screws. 
2.  Install the ISE shielding cover by tightening the screw. 
3.  Pull out the draining tube through the hole in the ISE shielding cover.(it is not 
indicated in figure) 
4.  Install the table panel and ISE unit door.