8-3
NOTE
Nebulization may cause fluctuation in the patient’s FiO2.
8.5 O2
The O2function means to deliver oxygen with concentration higher than the normal level.
When the O2key is pushed, the ventilator starts the O2 function. The O2indicator
lamp is illuminated and the screen prompt message filed displays the remaining O2 time.
O2 is active for a maximum of two minutes. When O2 is functioning, the [O2%]
parameter in the parameter setup quick key field displays the currently set oxygen
concentration.
When two minutes of O2 expires or the O2key is pushed again, the ventilator
terminates O2 function. Removing the breathing hoses during O2starts the suction
function.
NOTE
When O2 supply pressure is low, this function is not initiated, and is terminated
automatically if this function is initiated already.
8.6 Suction
The ventilator provides a suction procedure to help the patient to complete suction.
1. Push the Menu key and select [Procedure].
2. Select [Suction] to access the suction procedure screen and activate the suction function.
When the suction procedure is activated, the system performs O2function and
prompts [O2 Active. Please disconnect the patient after adequate O2 is
delivered.]. If the patient hoses are not disconnected within the specified time, the
suction procedure will end automatically.
After the patient hoses are disconnected, the system prompts [The patient is
disconnected. Please reconnect the patient after suction is completed.] and
stops ventilating the patient. In this case, manual sunction can be applied to the
patient.
After manual suction applied to the patient, re-connect the patient hoses. The
system performs O2function and prompts [O2Active. The patient is
reconnected.].
To stop an active suction procedure during O2, select [Exit] or push the O2key.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SynoVent E5 and is the answer not in the manual?

Mindray SynoVent E5 Specifications

General IconGeneral
BrandMindray
ModelSynoVent E5
CategoryMedical Equipment
LanguageEnglish

Related product manuals