E-2 
 
E.3 Ventilation Mode 
Ventilation mode  Factory default setting 
V-A/C 
TV  Adult: 7 kg/mL x IBW and 100 ml, whichever is greater; pediatric: 
7 kg/mL x IBW (round down) (when TV/f source is IBW or height 
& gender) 
Adult: 500 ml; pediatric: 50 ml (when TV/f source is patient type) 
O2% 40% 
Plimit 40 cmH2O 
PEEP 3 cmH2O 
△int.PEEP OFF 
Sigh OFF 
Tinsp  Adult: 1.7s; pediatric: 0.7s 
I:E 1:2 
Assist ON 
F-Trig  Adult: 2.0 L/min; pediatric: 1.0 L/min 
Flow  Adult: 20 L/min; pediatric: 8 L/min 
f  Adult: 12 bpm; pediatric: 29 bpm 
ATRC (only in non-NIV 
mode) 
OFF 
Tube diameter (only in 
non-NIV mode) 
8.0mm 
Compensation proportion 
(only in non-NIV mode) 
80% 
Compensation of expiration 
(only in non-NIV mode) 
ON 
P-A/C 
O2% 40% 
PEEP 3 cmH2O 
Pinsp 15 cmH2O 
Tinsp  Adult: 1.7s; pediatric: 0.7s 
I:E 1:2 
Tslope 0.2s 
Assist ON 
F-Trig  Adult: 2.0 L/min; pediatric: 1.0 L/min