E-1
E Factory Defaults
This chapter lists the most important factory default settings which are not user-adjustable.
When necessary, you can restore the factory default settings.
E.1 CO2 Module
CO2 module Factory default setting
Working mode Measure
Pump rate 100 mL/min
Humidity compensation OFF
CO2 calibration concentration
setting
3%
EtCO2 low limit Adult: 15 mmHg; Pediatric: 20 mmHg
EtCO2 high limit Adult/pediatric: 50 mmHg
FiCO2 high limit Adult/pediatric: 4 mmHg
E.2 Alarm
Alarm Factory default setting
Paw high limit 50 cmH2O
MV high limit
TV*f*1.5
MV low limit
TV*f*0.8
TVe high limit TV setting value x 2
ftot high limit OFF
Tapnea 15s
Alarm Volume 2