B-1 
B Product Specifications 
 
The ventilator is already integrated with expiratory volume monitor, pressure measurement 
device, and pressure release device. It has a built-in gas mixer and is equipped with alarm 
system, O
2
 monitor, and CO
2
 monitor, where: 
  The expiratory volume monitor, pressure measurement device, and pressure release 
device comply with IEC60601-2-12 . 
  The expiratory volume monitor, pressure measurement device, and pressure release 
device comply with IEC80601-2-12. 
  The built-in gas mixer complies with ISO11195. 
  The alarm system complies IEC 60601-1-8. 
  The O
2 
monitor complies with ISO21647. 
  The CO
2 
monitor complies with ISO21647. 
  The O2 monitor complies with IEC80601-2-55.   
  The CO2 monitor complies with IEC80601-2-55. 
B.1 Safety Specifications 
Type of protection against 
electric shock 
Class I equipment with internal electrical power supply. 
Where the integrity of the external protective earth (ground) in 
the installation or its conductors is in doubt, the equipment 
shall be operated from its internal electrical power supply 
(batteries). 
Degree of protection against 
electric shock 
BF, defibrillation-proof 
Operating mode  Continuous 
Degree of protection against 
hazards of explosion 
Ordinary equipment, without protection against explosion; not 
for use with flammable anaesthetics. 
Degree of protection against 
harmful ingress of water 
Degrees of protection provided by enclosures(IP 
Code)——IP21 
Protection Index according the EN 60529 standard: 
2: Protected against solid foreign objects of 12.5 mm diameter 
and greater 
1: Protected against vertically falling water drops 
Electrical connections between 
the equipment and the patient 
Non-electrical connections 
Equipment type  Mobile 
Disinfection  Steam autoclavable or other disinfection method.