EasyManuals Logo

Mindray SynoVent E5 User Manual

Mindray SynoVent E5
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #166 background imageLoading...
Page #166 background image
B-12
B.11 Compressor Specifications
Compressor Specifications
Input voltage 220 to 240 V 110 to 120 V
Input frequency 50/60Hz 60Hz
Input current 3A 6A
Output pressure range 300 to 450 kPa
Noise Less than 50dB (A)
Continuous flow 30 L/min at output pressure of 300kPa
Peak flow Greater than 180 L/min for more than 0.8 seconds at one
barometric pressure
Dew point Lower than the room temperature by 5 at flow of 30 L/min
B.12 O
2
Sensor Specifications
O2 sensor
Output 9 to13 mV at 210 hPa O2
Range 0 to 1500hPa O2
100% O2 signal deviation 100±1%
Resolution 1 hPa O2
Expected service life 1.5 x 106 % for measurement (20°C)
0.8 x 106 % for measurement (40°C)
Response time ((21% air to 100% O2) <15s
Linearity Linear 0-100% O2
Operating temperature range -20°C to +50°C
Temperature compensation ±2% of fluctuation at 0-40°C
Pressure range 50 to 200 kPa
Relative humidity 0 to 99%
100% O2 concentration output drift Over one year of typical value <5%
Material White ABS
Packaging Sealed package
Period of validity Not more than 13 months after unpacked (in compliance
with the conditions specified by the manufacturer)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SynoVent E5 and is the answer not in the manual?

Mindray SynoVent E5 Specifications

General IconGeneral
BrandMindray
ModelSynoVent E5
CategoryMedical Equipment
LanguageEnglish

Related product manuals