11-4 
 
11.4 Flow Sensor Calibration 
NOTE 
  Do not perform calibration while the unit is connected to a patient. 
  During calibration, do not operate the pneumatic parts. Do not move or press the 
breathing hoses especially. 
  Make sure that the system is Standby. If not, push the Standby key to enter 
standby screen. 
  It is recommended not to connect the humidifier to the ventilator before the 
calibration. 
 
Calibrate the flow sensor when the measured value has a great deviation or when the flow 
sensor is replaced. 
Follow these steps to calibrate the flow sensor: 
1.  Make sure that Air and O2 supplies are connected. 
2.  Connect the breathing hoses and insert the Y piece into the leak test plug to close the 
breathing circuit. 
3.  Push the Menu key. Select [Calibrate] and select [Flow]. Select [Calibrate] on the 
right side to start flow sensor calibration. The message [Calibrating] is displayed. 
4.  During the calibration, if you select [Stop], the ongoing calibration will stop and the 
message [Calibration Stopped! Calibration is unfinished.] is displayed. 
5.  After a successful calibration of both Air and O2 supplies, the screen shows 
[Calibration Completed!]. Otherwise, the message indicating calibration failure is 
displayed. In this case, you need to do the calibration again. 
 
NOTE 
  In case of calibration failure, check for sensor malfunctioning alarm and then 
troubleshoot it if there is. If it still fails or great measurement error occurs after 
calibration, replace the flow sensor and repeat the above operation. If the 
measurement error is still great, contact your service personnel or us.