B-11
B.10 CO
2
Module Specifications
CO
2
module
Measurement range and
accuracy
Measurement range Accuracy
0 to 40 mmHg ±2 mmHg
41 to 76 mmHg ±5% of reading
77 to 99 mmHg ±10% of reading
Resolution 1 mmHg
Rise time
<330ms@100mL/min
<400ms@70mL/min
Delay time
<3s@100mL/min
<3.5s@70mL/min
Measured by using neonatal water trap and 2.5-meter long
neonatal sampling line.
<5s@100mL/min
<6.5s@70mL/min
Measured by using adult water trap and 2.5-meter long adult
sampling line.
Total system response time
<3.5s@100mL/min
<4s@70mL/min
Measured by using neonatal water trap and 2.5-meter long
neonatal sampling line.
<5.5s@100mL/min
<7s@70mL/min
Measured by using adult water trap and 2.5-meter long adult
sampling line.
Pump rate
70 mL/min and 100 mL/min optional.
The flow control accuracy is ±15 of the set value or ±
15mL/min, whichever is greater.
Sidestream CO
2
alarm limits Range Step
EtCO
2
high limit (low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
low limit 0 to (high limit - 2) mmHg
FiCO
2
high limit 1 to 99 mmHg