B-2
B.2 Environmental Specifications
Main unit
Item Temperature (ºC) Relative humidity
(non-condensing)
Barometric
pressure (kPa)
Operating +10 to +40 15 to 95% 50 to 106
Storage and transport –20 to +60 (O2 sensor: –20 to +50) 10 to 95% 50 to 106
B.3 Power Requirements
External AC power supply
Input voltage 220 to 240 V 100 to 120 V
Input frequency 50/60 Hz 50/60 Hz
Input current 10A 15A
Fuse T10AH/250V T15AH/250V
Main unit power supply
Input voltage 100 to 240 V
Input frequency 50/60 Hz
Input current 1.6 to 0.8A
Fuse T3.15AH/250V
Auxiliary output power supply
Output voltage 220 to 240 V 100 to 120 V
Output frequency 50/60 Hz 50/60 Hz
Output current(outlet 1) 3.0A 6.0A
Output current(outlet 2) 3.0A 3.5A
Output current(outlet 3) 3.0A 3.5A
Fuse(outlet 1) T5AH/250V T10AH/250V
Fuse(outlet 2) T5AH/250V T5AH/250V
Fuse(outlet 3) T5AH/250V T5AH/250V
Internal battery
Number of batteries One or two
Battery type Lithium-ion battery
Rated battery voltage 11.1 VDC
Battery capacity 4500 mAh for a single battery
Overcurrent protection 8±2A
Time to shutdown
10 min at least (powered by new fully-charged batteries after the
first low battery alarm)
Battery run time
90 min (powered by one new fully-charged battery at 25℃ ambient
temperature);
180 min (powered by two new fully-charged batteries at 25℃
ambient temperature).