B-7
P0.1 -20 to 0 0.1 cmH
2
O
PEEPi 0 to 120 0.1 cmH
2
O
FiO
2
15 to 100 1 %
B.7 Ventilator Accuracy
Control accuracy
O
2
% ±3 vol.% or ±5% of setting, whichever is greater
TV ±10 mL or ±10% of setting, whichever is greater
f 1 bpm to 100 bpm: ±1 bpm
Other range: ±2% of setting
fSIMV ±1 bpm
Tinsp ±0.1s or ±10% of setting, whichever is greater
I: E
2: 1 to 1: 4: ±10% of setting
Other range: not defined.
Flow ±1 L/min or ±20% of setting, whichever is greater
Tslope ±0.2s
Plimit ±2.0 cmH
2
O or ±10% of setting, whichever is greater
PEEP ±2.0 cmH
2
O or ±10% of setting, whichever is greater
Pinsp ±2.0 cmH
2
O or ±10% of setting, whichever is greater
△Psupp ±2.0 cmH
2
O or ±10% of setting, whichever is greater
Phigh ±2.0 cmH
2
O or ±10% of setting, whichever is greater
Plow ±2.0 cmH
2
O or ±10% of setting, whichever is greater
Thigh ±0.2s or ±10% of setting, whichever is greater
Tlow ±0.2s or ±10% of setting, whichever is greater
Trigger ±1.0 cmH
2
O or ±10% of setting, whichever is greater or,
±1.0 L/min or ±20% of setting, whichever is greater
△int.PEEP ±2.0 cmH
2
O or ±20% of reading whichever is greater
Exp% ±10%
fapnea ±1 bpm
△Papnea ±2.0 cmH
2
O or ±10% of setting, whichever is greater
Monitoring accuracy
Ppeak
±(2 cmH
2
O + 4 % of the actual reading)
Pplat
Pmean
PEEP