B-8
TVi
0 ml~100 ml: ±10 ml + 3% of reading
100 ml~4000 m: ±5 ml + 8% of reading
±15% of reading or ±15 mL, whichever is greater (NIV)
TVe
TVe spn
MV
±8% of reading or ±0.3 L/min, whichever is greater
MVspn
MVleak
ftot
±5% of reading or ±1bpm, whichever is greater
fmand
fspn
Rinsp
0 to 50: ±10 cmH
2
O/(L/s)
Other range: not defined.
Rexp
Cstat 0 to 100: ±10 mL/cmH
2
O or ±20% of the displayed value,
whichever is greater
Other range: not defined.
Cdyn
RSBI
0 to 1000: ±20 1/(min·L)or ±15% of the displayed value, whichever
is greater
Other range: not defined.
WOB No declaration
NIF ±(2 cmH
2
O + 4 % of the actual reading)
P0.1 ±(2 cmH
2
O + 4 % of the actual reading)
PEEPi No declaration
FiO
2
21% to 100%: ±3 vol.%
Other range: not defined.
Oxygen concentration
controlling responded time
the responded time of the oxygen concentration in the delivered
volume to change from a volume fraction of 21 % to 90 %:
when reusable adult pipeline is used, TV=500ml, f=10bpm,I:E=1:2,
40 s ;
when reusable pediatric pipeline is used, TV=150ml, f=20bpm,
I:E=1:2, 60 s ;
when reusable infant pipeline is used, TV=30ml, f=30bpm,
I:E=1:2, 40 s .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SynoVent E5 and is the answer not in the manual?

Mindray SynoVent E5 Specifications

General IconGeneral
BrandMindray
ModelSynoVent E5
CategoryMedical Equipment
LanguageEnglish

Related product manuals