C-1
C EMC
SynoVent E5 ventilator is in compliance with IEC 60601-1-2 for EMC.
The essential performance verified during the immunity testing comprised of V
del
control
accuracy, V
del
monitoring accuracy, monitoring of the airway pressure, CO2 Accuracy and O2
Accuracy.
NOTE
Using accessories other than those specified may result in increased
electromagnetic emission or decreased electromagnetic immunity of the equipment.
The ventilator or its components should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the ventilator or its
components should be observed to verify normal operation in the configuration in
which it will be used.
The ventilator needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
Other devices may interfere with this equipment even though they meet the
requirements of CISPR.
When the input signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
Use of portable or mobile communications devices can degrade the performance of
the equipment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SynoVent E5 and is the answer not in the manual?

Mindray SynoVent E5 Specifications

General IconGeneral
BrandMindray
ModelSynoVent E5
CategoryMedical Equipment
LanguageEnglish

Related product manuals