Air Supply 
Pressure Low 
H The supply gas pressure is low. 
Check the status of supply gas. If the alarm continues 
although supply gas is normal, contact your service 
personnel. 
O2 Supply 
Pressure Low 
H
No Gas Supply 
Pressure 
H Air supply pressure low and O2 supply pressure low occur 
every second simultaneously. 
1. Make sure that the patient is in safe ventilation status. 
2. Check the actual pressures of Air and O2 supplies. 
3. Check that Air and O2 supply devices and connections 
are in good condition. 
4. Check Air and O2 supply pressure switches. 
Airway Leak?  L The airway is leaky. Check the airway for leakage and 
restore it to good condition. 
Airway 
Obstructed? 
H The airway is obstructed. 
Check the airway for leakage and restore it to good 
condition. 
Tube 
Disconnected? 
H The tube is disconnected. 
Restore tube connection. 
TV Not Achieved  L Tidal volume is less than the set tidal volume by over 
70%. 
Check ventilation setting parameters to see if the 
ventilation volume is too large. Check if the supply gases 
are normal. Check if the expiratory flow sensor has a 
deviation and perform zero calibration. Check the hoses 
for leakage. 
Pinsp Not 
Achieved 
L Inspiratory pressure is less than the set inspiratory pressure 
by over 70%. 
Check ventilation setting parameters to see if the 
inspiratory pressure is too large. Check if TV high alarm 
limit is set too low. Check if the supply gases are normal. 
Check the hoses for leakage. 
Sustained Airway 
Pressure 
H The airway pressure measured by any pressure sensor is 
greater than or equal to the set PEEP+15 cmH2O for 15s 
continuously. 
Check the breathing circuit for occlusion. Check if the 
expiratory time is too short. 
PEEP Too High  H The measured PEEP exceeds PEEP+5 cmH2O within any 
fully mechanical ventilation cycle. 
Check the breathing circuit for occlusion. Check if the 
expiratory time is too short. 
Tinsp too Long  L In PSV mode, the gas delivery time exceeds 4.5s for adult 
and 1.5s for pediatric for three continuous cycles. This 
alarm is not triggered if pressure sensor or flow sensor is 
faulty.