2-6
3. Alarm indicator
The alarm indicator is lit when the compressor internal temperature is abnormally high.
In this case, the compressor may shut off at any time and stop delivering gas.
4. Pressure gauge
The pressure gauge indicates the air pressure at the compressed air outlet.
5. Compressed air outlet
6. Central pipeline gas supply inlet
7. Power switch
8. Mains power inlet (with fixing pressure plate)
9. Air intake vent (with dust filter)
10. Hourmeter
The hourmeter indicates the accumulated running time of the compressor (not including
the accumulated running time when the central pipeline gas supply is applied)
NOTE
Burn-in is required for the compressor before delivery. The reading indicated by
the compressor hourmeter shall be less than 150 hours at the time of delivery.