20 User Guide EN Dansensor® Lippke® 5000
P/N 380419-C
05/2022
How does it work?
1. Adjust the restraining height - see "Adjusting the restraining height (ASTM)" on page 21 for
details.
2. Open the restraining plates by sliding them away from each other.
3. Prepare the package with a septum and place it on the base plate .
4. Move the test head downwards to penetrate the package making sure that the
needle
is fully seated onto the septum .
5. Close the restraining plates . Lock the test head in the current position taking into
consideration that it should be allowed to move upwards as the package inflates.
If, for some reason, you want to restrict the test head movement, you can do this by means
of the knob
.
6. Perform the test.