Envelope drawing mode is always opposite the selected Live A-Scan drawing mode.
X-Axis Grid Mode
Sets the horizontal X-axis grid display mode
Off: no grid
Standard: 10 divisions spaced evenly and labeled 1 to 10
Soundpath: five divisions spaced evenly and labeled with corresponding
sound path values
Leg: up to four divisions representing half-skip distances in angle beam
inspection mode, based on test piece thickness value, and labeled L1 to L4
Y-Axis Grid
Sets the vertical Y-axis grid display mode to 100 % or 110 % full screen height
Baseline (Baseline Break)
Modifies the appearance of the A-scan in full wave rectified mode. When
active, the instrument locates all zero cross points in the RF waveform and
pulls the full wave rectified A-scan to the baseline. This feature helps to see
small defects that are close to the back surface of the test piece, especially at
large ranges.
On: Baseline Break enabled
Off: Baseline Break disabled
VGA Output
Controls the output to the VGA connector.
On
Off
Brightness
Used to adjust the screen brightness or by choosing one of the preset values
(0 %, 25 %, 50 %, 75 %, or 100 %).
Color Scheme
Sets the overall instrument color scheme
Default: default multi color display
Classic: EPOCH 600 color scheme
Outdoor W: white background, black text