DMTA-10055-01EN, Rev. A, February 2015
Chapter 9
154
Figure 9-7 Entering the Zero Cal thickness value
11. Adjust the value to match the known thickness of the gated indication (0.100 in. in
this example), and then press Continue (P3 key) to continue to the second
calibration step (see Figure 9-8 on page 155). The thickness value used in this part
of the calibration will be stored for reference in the Cal-Zero parameter box.
If you need to exit without acquiring calibration data, press Cancel (P1 key).