DMTA-10055-01EN, Rev. A, February 2015
Managing Special Waveform Functions
117
Figure 6-2 Peak memory signal envelope example
This function is useful when it is necessary to find the peak from an indication during
an angle beam inspection.
The peak memory function cannot be activated in the unrectified Rcvr > Rect = RF
mode.
To activate peak memory function
1. Press PEAK MEM.
The symbol appears in the flag area to indicate that the function is active.
2. Scan over a reflector to acquire the echo envelope.
3. Press PEAK MEM again to turn off the peak memory function.
Peak memory flag
Peak memory signal envelope
(blue area)
Current live waveform (green
line)