EasyManua.ls Logo

Olympus EPOCH 650 - Connectors

Olympus EPOCH 650
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10055-01EN, Rev. A, February 2015
Hardware Overview
33
Figure 1-15 Alarm indicators
The alarm indicators illuminate in red when the corresponding gate alarm is
triggered. See “Gate Alarms” on page 133 for information on gate alarms.
1.2 Connectors
The EPOCH 650 is equipped with connectors to allow the necessary and optional
connections.
1.2.1 Transducer Connectors
The EPOCH 650 instrument is supplied with either BNC or LEMO 01 series
transducer connectors. The type of transducer connector is chosen at the time of order.
If necessary, it is possible to change the type of transducer connection at an authorized
Olympus service center for a small charge. Selection of these transducer connections
is based on operator preference. Both the BNC and LEMO 01 series connectors
available are rated to IP67 for use in most inspection environments. The EPOCH 650
is illustrated with BNC connectors in this document.
The transducer connectors are located at the top of the instrument on the left. The two
connectors are easily accessible from the front of the instrument (see Figure 1-16 on
page 34).
Gate 1 alarm indicator
Gate 2 alarm indicator
Interface gate alarm indicator

Table of Contents

Other manuals for Olympus EPOCH 650

Related product manuals