DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
242
Figure 11-14 DGS/AVG setup page
The DGS/AVG setup page allows you to define the exact probe being used for the
inspection, and to set up the DGS/AVG curve to be drawn. There are several
adjustments that can be made from this setup page:
DGS/AVG
Activate/Deactivate the DGS/AVG function
Probe Type
Selects the type of probe to be used (Straight Beam, Angle Beam, or Dual
Probe).
Probe Name
Selects the actual probe to be used.
Reflector Type
Defines the type of reflector to be used to acquire a reference indication to
build the DGS/AVG curve. For straight beam and dual probes, the available
reflectors are:
Back wall
Side-drilled hole (SDH)
Flat-bottom hole (FBH)