DMTA-10055-01EN, Rev. A, February 2015
Chapter 4
76
4.3 Direct-Access Parameters
Direct-access parameters always appear at the top right corner of the display window.
This allows you to view these parameters from the main screen regardless of which
menu or parameter you are currently accessing. These parameters include Gain,
Range, Delay, and Gates.
To select and view the direct-access parameters
• Press the dB direct-access key to select the Gain parameter.
• Press the RANGE direct-access key to select the Range parameter.
• Press the GATES direct-access key to select the Gate parameters. The G1Start
parameter replaces the Delay parameter.
•Press the 2
ND
F (DELAY) direct-access key to select the Delay parameter.
Once selected, the button turns green (see Figure 4-5 on page 76). Turn the adjustment
knob or press the Up and Down arrow keys to edit the value. For additional
information on gate adjustment, see “Quickly Adjusting Basic Gate Parameters” on
page 125.
Figure 4-5 Example of the Gain, Range, and Delay direct-access parameters
When you are using the gain, range, or delay direct access keys, a set of preset values
display over the parameter keys at the bottom of the screen (see Figure 4-6 on
page 77). These preset values allow you to jump to a particular setting with a single
button press, and can be customized (see “Editable Parameters Page” on page 102).
Press RANGE to select
Press 2
ND
F, (DELAY) to select. Press
GATES to return to gate parameter
Press
dB to select