6.2 Peak Memory
The peak memory function enables the display to capture and store on the screen the
amplitude of each A-scan acquisition. The display updates each pixel if a signal of
greater amplitude is acquired. When you scan the transducer over a reflector, the
signal envelope (echo dynamic as a function of transducer position) is held on the
screen as a green line (see Figure 6-2 on page 117). In addition, the current, live
waveform is displayed at the appropriate place within the signal envelope.