DMTA-10055-01EN, Rev. A, February 2015
Chapter 3
64
Figure 3-14 The Cal-Zero value
8. Select Continue to accept the adjusted value, and then move on to the second
calibration step.
9. Couple the transducer to the thick step of material.
10. Position gate 1 around the resulting indication on screen.
11. Bring the indication to 80 % full screen height using the 2
ND
F, (AUTO XX%) keys
(see Figure 3-15 on page 65).