DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
256
Figure 11-21 Completed API 5UE sizing
11.5.1 Option Activation and Setup
If the API 5UE software option has been licensed on the instrument, it can be activated
at any time during standard operation.
To activate the API 5UE software option
1. Select API5UE > Setup.
2. In the API5UE menu, select API5UE = On (see Figure 11-22 on page 257).
3. Press NEXT, and then enter the known height of the calibration reference notch in
the Ref. Depth box.
4. Press to return to the live screen.