13. Once a steady reading is achieved, choose Auto Cal > Cal-Vel.
The screen freezes and the Cal-Vel box appears (see Figure 9-15 on page 164.
Figure 9-15 Entering the Velocity Cal thickness value
14. Adjust the value to match the known thickness of the gated indication (0.300 in. in
this example), and then choose Done to complete the calibration process.
If for any reason you need to exit without acquiring calibration data, press Cancel.