DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
262
Figure 11-25 Storing the A
max
point
7. Move the transducer forward over the notch until the peak drops to 40 % FSH on
the leading edge of the signal.
8. Press RefT1 (P2) to collect the 6 dB drop position of the leading peak and assign
this value as T
1
(see Figure 11-26 on page 263).