DMTA-10055-01EN, Rev. A, February 2015
Software Features and Options
291
2. Ensure that the B-scan is active (see “Activating the B-Scan” on page 286).
3. View the Contents of the file ID with the saved B-scan (see “Viewing Setup and
Waveform Data for Saved IDs in a File” on page 199).
A vertical cursor on the B-scan image indicates the position of the saved
compressed A-scan. The vertical cursor is controlled by buttons that are displayed
above the parameter keys (P4 - P7) when a B-scan file is active (see Figure 11-47
on page 291):
Left arrow
(P4 key)
Moves the vertical cursor one (1) increment to the left for review of individual
B-scan data points.
Right arrow (P5 key)
Moves the vertical cursor one (1) increment to the right for review of
individual B-scan data points.
Left double arrow (P6 key)
Moves the vertical cursor ten (10) increments to the left for rapid review of
individual B-scan data points.
Right double arrow (P7 key)
Moves the vertical cursor ten (10) increments to the right for rapid review of
individual B-scan data points.