DMTA-10055-01EN, Rev. A, February 2015
Chapter 13
302
Table 16 Environmental ratings specifications
Parameter Va lu e
IP rating Designed to meet requirements of IP66 (knob
configuration) or IP67 (navigation pad configuration)
Explosive atmosphere MIL-STD-810F Procedure 1, NFPA 70E, Section 500,
Class 1, Div. 2, Group D
Shock tested IEC 60068-2-27, 60 g, 6 µs H.S., multiple axes, 18 total
Vibration tested Sine vibration, IEC 60068-2-6, 50 Hz to 150 Hz at
0.762 mm. (0.03 in.). DA or 2 g, 20 sweep cycles
Operating temperature –10 °C to 50 °C (–14 °F to 122 °F)
Battery storage temperature 0 °C to 50 °C (32 °F to 122 °F)