Olympus EPOCH 650 - Figure 9-1 Example of a Gated Signal for Zero Calibration

Olympus EPOCH 650
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10055-01EN, Rev. A, February 2015
Calibration
147
6. Press dB, and then adjust the gain value so that the echo amplitude is at
approximately 80 %.
The AUTO XX% feature automatically adjusts the gain to set the gated echo
amplitude to XX% of full screen height (the default XX value is 80 %). To activate this
feature, press 2
ND
F, (AUTO XX%).
A thickness measurement reading appears in large text above the A-scan (see
Figure 9-1 on page 147).
Figure 9-1 Example of a gated signal for zero calibration
7. Once a steady reading is achieved, choose Auto Cal > Cal-Zero.
The screen freezes and the Cal-Zero box appears (see Figure 9-2 on page 148).
Thickness measurement