DMTA-10055-01EN, Rev. A, February 2015
Software Features and Options
263
Figure 11-26 Storing the T
1
point
9. Move the transducer backward over the notch until the peak drops rises to 80 %
and then falls back to 40 % FSH on the trailing edge of the signal.
10. Press RefT2 (P3) to collect the 6 dB drop position of the trailing peak and assign
this value as T
2
(see Figure 11-27 on page 264).