Loading...
DMTA-10055-01EN, Rev. A, February 2015
Software Interface
93
Gate 1 Depth to Reflector Depth to reflector in gate 1. Used with
Angle.
Gate 2 Depth to Reflector Depth to reflector in gate 2. Used with
Angle.
Gate 1 Surface Distance Horizontal distance to reflector in gate 1.
Used with Angle.
Gate 2 Surface Distance Horizontal distance to reflector in gate 2.
Used with Angle.
Gate 1 Surface Dist – x Val Horizontal distance minus X-Value
(distance from beam index point to front of
wedge) in gate 1. Used with Angle.
Gate 2 Surface Dist – x Val Horizontal distance minus X-value
(distance from beam index point to front of
wedge) in gate 2. Used with Angle.
Gate 1 Minimum Depth Minimum depth in gate 1. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 2 Minimum Depth Minimum depth in gate 2. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 1 Maximum Depth Maximum depth in gate 1. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 2 Maximum Depth Maximum depth in gate 2. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 1 Current Amplitude Amplitude measurement in gate 1.
Displays as % of full-screen height (FSH).
Gate 2 Current Amplitude Amplitude measurement in gate 2.
Displays as % of full-screen height (FSH).
Gate 1 Maximum Amplitude
Maximum amplitude in gate 1. Resets on
gate adjustment and on most
pulser/receiver adjustments.
Table 13 Available measurement readings (continued)
Icon Measurement reading Description

Table of Contents