Slide Scanner Repair Manual Diagnostics and Troubleshooting
115
Table 5-5. Scanner Tester Panel
Item Box Description
Scanner SN Used to enter the serial number (ID) of the slide scanner to be
tested.
Station Name If applicable, used to enter the test station name.
CCD Configuration Allows the user to update or leave unchanged, the CCD
configuration. This selection is used when the CCD is
replaced with a newer generation unit.
CCD Config Data After a test cycle is completed, the scanner test panel reports
the type of CCD used in the slide scanner. Type 0 or 1.
GMT Test Sequence Lists the image quality tests that are sequenced when the
Geometric (GMT) test target is being used.
1 = Do test -Allows the user to select all or a particular
image quality test. The tester default is all
tests selected.
One (1) indicates that the particular image
quality test is selected.
Zero (0) indicates that the particular image
quality test is not selected.
P/F (Pass/Fail)-Indicates whether or not a selected image
quality test passed (P) or failed (F).
CT Test Sequence Lists the image quality tests that are sequenced when the Color
(CT) test target is being used.
1 = Do test -Allows the user to select all or a particular
image quality test. The tester default is all
tests selected.
One (1) indicates that the particular image
quality test is selected.
Zero (0) indicates that the particular image
quality test is not selected.