Slide Scanner Repair Manual Scanner Overview
29
Check and adjust, if
necessary:
• Base line integration time
to ensure a minimum signal
and mitigation of saturation
• Analog gain and offset for
base line neutral settings
• Digital uniformity
offset calibration
• Digital uniformity
gain calibration
• Bad pixel identification
• Drift test
Figure 1-8. Calibration simplified flow diagram
Calibration
Completed
Ready LED
Lights Steady
Fail
Run LabVIEW Test
(See Section 5 for Details)
Ready LED
Continually Blinks
Start
First Scan
Start
Subsequent Scans
Calibrate
Check
Drift in Lamp
or
Front End
Fail
Pass
SS35/SS35 ES
Calibration Routine
Pass
Scan
No Scan