Slide Scanner Repair Manual Calibration and Adjustments
Figure 3-8. Expanded red waveform
4. Manually rotate the lead screw of the chassis assembly until the focus/magnification target
is properly positioned (Figure 3-9) in the viewing area of the slide scanner.
Note: Figure 3-8 displays an expanded Red waveform after the target is inserted.
The oscilloscope time base settings are adjusted to view only the Red portion
of the VSP output.
Figure 3-9. Focus/magnification target
5. Examine the Red channel output of image produced by the target. Figure 3-10 shows
the left side of the target (Figure 3-9, points 1, 2, 3) and the right side of the target
(Figure 3-9, points 4, 5, 6).
The left and right sides of the target are intensified and displayed so that we can see if
they are of equal amplitude. When 1 & 3 equals 4& 6 the slide scanner is focused
correctly.
Viewing Area
123 456
t
Left Side
Right Side