Send stable measurement result in basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- command carried out, immediate response: mass value in basic
measuring unit
Response format:
1
2-3
4
5
6
7-15
16
17
18
19
20
21
S space
stability
marker
space character mass space unit CR LF
Example:
S CR LF
- command sent from a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
- command carried out, immediate
response: mass value in current
measuring unit.
where: _ - space
Immediately send measurement result in basic measuring unit
Format: SI CR LF
Response options:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S I space
stability
marker
space character mass space unit CR LF
Example:
S I CR LF
- command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
- command carried out, immediate
response: mass value in basic measuring
unit
where: _ - space
Send stable measurement result in current measuring unit
Format: SU CR LF
Response options:
SU_A CR LF
- command understood and in progress
SU_E CR LF
- time limit exceeded while waiting for stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- command carried out, immediate response: mass value in basic
measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21

Table of Contents