- 107 -
S U space
stability
marker
space character mass space unit CR LF
Example:
S U CR LF
- command sent from a computer
S U _ A CR LF
- command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
- command carried out, response: mass
value in current measuring unit.
where: _ - space
Immediately send measurement result in current measuring unit
Format: SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
Example:
S U I CR LF
- command sent from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate
response: mass value in current
measuring unit.
where: _ - space
Switch on continuous transmission in basic measuring unit
Format: C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- command carried out, immediate response: mass value in basic
measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S I space
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress

Table of Contents