Operation not possible for verified moisture analyzers. 
Give moisture analyzer serial number 
- command understood, response: serial number 
- command understood but not accessible at this moment 
 – moisture analyzer serial number, inserted in between inverted commas 
NB CR LF - give serial number 
NB_A_”1234567” - weighing device serial number –-”1234567”  
Lock moisture analyzer keypad 
- command understood but not accessible at this moment  
Command locks the moisture analyzer keypad (proximity sensors, touch panel) until the 
moment of turning the moisture analyzer off or until sending K0 command. 
Unlock moisture analyzer keypad 
- command understood but not accessible at this moment  
Give available working modes 
Command returns accessible working modes. 
n_”Mode name” <CR><LF> 
: 
n_”Mode name” <CR><LF> 
 command carried out, response: accessible working 
- command understood but not accessible at this moment 
Mode name – parameter, working mode name, inserted in between inverted comas. The 
name takes form given on a particular moisture analyzer display, it is provided in a currently 
n - parameter, decimal value determining working mode number. 
Working modes numbering is identical for each kind of weighing instrument. The numbers 
are assigned to working modes names. 
Some moisture analyzers give only the number as a response.