- 109 -
SM_OK CR LF
- command carried out
SM_I CR LF
- command understood but not accessible at this moment (e.g. mode
other than PARTS COUNTING)
ES CR LF
- command not recognised (mass format incorrect)
Set target mass value (e.g. DOSING mode)
Format: TV_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
TV_OK CR LF
- command carried out
TV_I CR LF
- command understood but not accessible at this moment (e.g. mode
other than DOSING)
ES CR LF
- command not recognised (mass format incorrect)
Set reference mass value (e.g. for PERCENT WEIGHING)
Format: RM_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
RM_OK CR LF
- command carried out
RM_I CR LF
- command understood but not accessible at this moment (e.g. mode
other than PERCENT WEIGHING)
ES CR LF
- command not recognised (mass format incorrect)
Value release
Format: SS CR LF
Response options:
SS_OK CR LF
- command understood and in progress
Command’s function is similar to function of PRINT key located on the instrument panel.
Internal adjustment
Format: IC CR LF
Response options:
IC_A CR LF
- command understood and in progress
IC_D CR LF
- adjustment completed
IC_A CR LF
- command understood and in progress
IC_E CR LF
- time limit exceeded while waiting for stable measurement result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal adjustment of the moisture analyzer
Format: IC1 CR LF
Response options:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation not possible, e.g. in case of verified moisture analyzer
IC1_OK CR LF
- command carried out
Operation not possible for verified moisture analyzers.
For non-verified moisture analyzers the command disables internal adjustment until it is
enabled via IC0 command or until the moisture analyzer is turned off. The command does
not change module settings that concern factors determining adjustment process start.
Enable automatic internal adjustment of the moisture analyzer
Format: IC0 CR LF
Response options:
IC0_I CR LF
- command understood but not accessible at this moment